Gas-Solid Headspace-GC Determination of Epoxyethane in Disposable Sanitary Products
摘 要
应用气固顶空-气相色谱法测定卫生护垫、纸尿裤、卫生巾等一次性卫生用品中残留环氧乙烷的含量。样品经剪碎后,在60 ℃下加热30 min,并取顶空气体进气相色谱以氢火焰离子化检测器进行检测。以基体匹配校正法有效消除基质效应。在卫生护垫、纸尿裤、卫生巾基质中,环氧乙烷的质量均在2~120 μg范围内与其峰面积呈线性关系,检出限(3S/N)均为3 μg·g-1。方法应用于一次性卫生用品的测定,加标回收率在89.5%~105%之间,测定值的相对标准偏差(n=6)在1.3%~2.9%之间。
Abstract
Gas-solid headspace-GC was applied to the determination of residual amount of epoxyethane in disposable sanitary products including sanitary pad, paper diaper, sanitary towel and so on. After being cut to pieces, the sample was heated to 60 ℃ for 30 mins for equilibrium. Then the headspace gas was extracted and injected into gas chromatograph for analysis by flame ionization detector. The matrix effect was eliminated by matrix-matching calibration. In the matrixes of sanitary pad, paper diaper and sanitary towel, linear relationships between values of peak area and mass of epoxyethane were kept in the same range of 2-120 μg with the same detection limit (3S/N) of 3 μg·g-1. The proposed method was applied to the analysis of samples of disposable sanitary products, giving values of recovery and RSD′s (n=6) in the ranges of 89.5%-105% and 1.3%-2.9% respectively.
中图分类号 O657.7
所属栏目 工作简报
基金项目 质检公益性行业科研专项(2012104020);一次性使用卫生用品国内外卫生检测标准的分析研究(HK013-2012)
收稿日期 2014/1/28
修改稿日期
网络出版日期
作者单位点击查看
备注马明(1980-),男,上海人,工程师,主要从事仪器分析工作。
引用该论文: MA Ming,SHEN Jie,YAO Li-fang,ZHOU Yu-yan. Gas-Solid Headspace-GC Determination of Epoxyethane in Disposable Sanitary Products[J]. Physical Testing and Chemical Analysis part B:Chemical Analysis, 2014, 50(9): 1117~1120
马明,沈劼,姚丽芳,周宇艳. 气固顶空-气相色谱法测定一次性卫生用品中的环氧乙烷[J]. 理化检验-化学分册, 2014, 50(9): 1117~1120
共有人对该论文发表了看法,其中:
人认为该论文很差
人认为该论文较差
人认为该论文一般
人认为该论文较好
人认为该论文很好
参考文献
【1】杨必勇,江生,张小松.气相色谱法测定医用棉签中环氧乙烷残留量[J].中国药业, 2006,15(20):30-31.
【2】李丹,王贞媛,夏顺宁.顶空毛细管气相色谱法测定医疗器械中残留环氧乙烷的残留量[J].中国医疗器械杂志, 2007,31(3):201-202.
【3】GB 15979-2002一次性使用卫生用品卫生标准[S].
【4】刘园园,黄海萍,陈岳蓉.顶空气相色谱法测定卫生巾中环氧乙烷残留量[J].中国药师, 2013,16(9):1338-1340.
【5】熊冬生,林福兰.气相色谱法测定医用高分子材料硅凝胶中环氧乙烷的残留[J].光谱实验室, 2004,21(1):146-149.
【6】王昊阳,郭寅龙,张正行,等.顶空-气相色谱法进展[J].分析测试技术与仪器, 2003,9(3):129-135.
【7】BRUNO K, LESLIE S E. Stactic headspace-gas chromatography: theory and practice (second edition)[M]. New Jersey: John Wiley and Sons Inc, 2006:34-35.
【2】李丹,王贞媛,夏顺宁.顶空毛细管气相色谱法测定医疗器械中残留环氧乙烷的残留量[J].中国医疗器械杂志, 2007,31(3):201-202.
【3】GB 15979-2002一次性使用卫生用品卫生标准[S].
【4】刘园园,黄海萍,陈岳蓉.顶空气相色谱法测定卫生巾中环氧乙烷残留量[J].中国药师, 2013,16(9):1338-1340.
【5】熊冬生,林福兰.气相色谱法测定医用高分子材料硅凝胶中环氧乙烷的残留[J].光谱实验室, 2004,21(1):146-149.
【6】王昊阳,郭寅龙,张正行,等.顶空-气相色谱法进展[J].分析测试技术与仪器, 2003,9(3):129-135.
【7】BRUNO K, LESLIE S E. Stactic headspace-gas chromatography: theory and practice (second edition)[M]. New Jersey: John Wiley and Sons Inc, 2006:34-35.
相关信息