提出了电感耦合等离子体原子发射光谱法测定防锈铝合金中痕量钠的方法。以盐酸(1+1)溶液和过氧化氢溶解样品,选择波长588.995 nm为钠的分析线。方法的检出限(3S/N)为0.028 mg·L-1。方法用于测定防锈铝合金中痕量钠(wNa=0.000 14%~0.000 73%),加标回收率在91.4%~113.3%之间,相对标准偏差(n=6)在2.6%~5.5%之间。
Abstract
An ICP-AES method for the determination of sodium in anticorrosive aluminum alloy was proposed. The sample was dissolved with HCl(1+1) and H2O2. The wavelength of 588.995 nm was selected as analytical line for the determination of sodium. Detection limit (3S/N) of the method found was 0.028 mg·L-1. The proposed method was applied to the analysis of trace amounts of sodium in anticorrosive aluminum alloy (wNa=0.000 14%-0.000 73%), giving values of recovery obtained by standard addition method in the range of 91.4%-113.3% and RSD′s (n=6) in the range of 2.6%-5.5%.
中图分类号 O657.31
所属栏目 工作简报
基金项目
收稿日期 2011/3/9
修改稿日期
网络出版日期
作者单位点击查看
备注李芬(1963-),女,四川隆昌人,高级工程师,理学学士,主要从事光谱分析和化学分析的检测和研究工作。
引用该论文: LI Fen,LI Qi-hua,ZHOU Xi-lin,MA Ying,FAN Chao-ying. ICP-AES Determination of Trace Amounts of Sodium in Anticorrosive Aluminum Alloy[J]. Physical Testing and Chemical Analysis part B:Chemical Analysis, 2012, 48(4): 465~466
李芬,李启华,周西林,马英,樊朝英. 电感耦合等离子体原子发射光谱法测定防锈铝合金中痕量钠[J]. 理化检验-化学分册, 2012, 48(4): 465~466
被引情况:
【1】余锦,李启华,刘菊英,樊朝英,马英, "电感耦合等离子体原子发射光谱法测定钨合金中镍、铁、钴和锰",理化检验-化学分册 50, 1131-1134(2014)
共有人对该论文发表了看法,其中:
人认为该论文很差
人认为该论文较差
人认为该论文一般
人认为该论文较好
人认为该论文很好
参考文献
【1】王滨滨.高镁铝合金的钠脆性[J].黑龙江冶金, 2009,29(3):12-14.
【2】张素华,李庆华,郭至修.离子选择电极法测定高硅防锈铝合金中的钠[J].冶金分析, 1984,4(4):64-64.
【3】GB/T 13748.17-2005镁及镁合金化学分析方法 钾含量和钠含量的测定 火焰原子吸收光谱法[S].
【4】辛仁轩.等离子体发射光谱分析[M].北京:化学工业出版社, 2005:9.
【5】陈必友,李启华.工厂分析化验手册[M].2版.北京:化学工业出版社, 2009:550.