EVALUATION OF UNCERTAINTY OF MEASUREMENT IN THE FAAS DETERMINATION OF SILVER IN AQUEOUS SOLUTION
摘 要
分析了影响原子吸收光谱法(AAS)测定水溶液中银的主要因素,并对其所引起的不确定度分量进行评估.对含银0.10-2.00 mg·L-1的水溶液,置信水平95%时,测量结果的扩展不确定度为0.020 mg·L-1,有效自由度为50.
Abstract
The main factors,affecting the accuracy of the FAAS determination of silver in aqueous solution were discussed,and the uncertainties of measurements caused by these factors in the determination were evaluated.An expanded uncertainty of 0.020 mg·L-1,at 95% confidencel level and with an effective degree of freedom of 50,was obtained in the determination of 0.10 to 2.00 mg·L-1 of silver.
中图分类号 O657.31
所属栏目 工作简报
基金项目
收稿日期 2004/5/25
修改稿日期
网络出版日期
作者单位点击查看
备注陈少鸿(1957-),女,福建南安人,高级工程师,主要从事无机元素分析工作.
引用该论文: CHEN Shao-hong,ZHU Li-hui. EVALUATION OF UNCERTAINTY OF MEASUREMENT IN THE FAAS DETERMINATION OF SILVER IN AQUEOUS SOLUTION[J]. Physical Testing and Chemical Analysis part B:Chemical Analysis, 2006, 42(5): 385~387
陈少鸿,朱丽辉. 原子吸收光谱法测定水溶液中银不确定度的评定[J]. 理化检验-化学分册, 2006, 42(5): 385~387
被引情况:
【1】彭良敏,支河,梁翊,钟冬莲, "分析方法不确定度评定的通用数学模型",理化检验-化学分册 45, 558-560(2009)
共有人对该论文发表了看法,其中:
人认为该论文很差
人认为该论文较差
人认为该论文一般
人认为该论文较好
人认为该论文很好
参考文献
【1】JJF1059-1999,测量不确定度评定与表示[S].
【2】李慎安.测量不确定度表达百问[M].北京:中国计量出版社,2001.79-93.
【3】JJG 196-1990,常用玻璃量器检定规程[S].
【4】刘珍.化验员读本[M].北京:化学工业出版社,2000.463-464.
【2】李慎安.测量不确定度表达百问[M].北京:中国计量出版社,2001.79-93.
【3】JJG 196-1990,常用玻璃量器检定规程[S].
【4】刘珍.化验员读本[M].北京:化学工业出版社,2000.463-464.
相关信息