APPLICATION OF PVC-MEMBRANE IODIDE ION SELECTIVE ELECTRODE TO THE ANALYSIS OF SYSTEM CO-EXISTING OF BROMIDE AND IODIDE IONS
摘 要
介绍了聚氯乙烯膜碘离子选择电极(PVC-I- ISE)的制备方法,电极对碘离子的能斯特响应斜率为59 mV/pI,响应的线性范围为4.0×10-5~1.0×10-2mol·L-1,检出限为1.6×10-5mol·L-1,低于用银ISE的检出限,电极的使用寿命为一个月以上.将制备的碘离子选择电极作为指示电极用于对溴、碘混合离子体系进行连续分别电位滴定,并与以银电极为指示电极的电位滴定方法进行了比较.结果表明用PVC-I--ISE作指示电极,可在I--Br-共存的混合溶液中进行两离子的连续电位滴定,而用银-ISE时则不能.
Abstract
Preparation of PVC-membrane iodide ISE was described and its electrochemical behaviors,especially its use as an indicator electrode in the potentiometric titration of iodide or bromide in the presence of each other,were studied.Linear response of this electrode was found in the range of 4.0×10-5-1.0×10-2mol·L-1 of iodide ion,with Nernst response slope of 59 mV/pI.Life of this I--ISE was found to be more than a month.As shown by experimental results,bromide and iodide in a mixture were titrated satisfactorily in succession by using the PVC-I--ISE,but it was impossible by using a silver-ISE.Besides,the detection limit of the PVC-I--ISE was found to be 1.6×10-5mol·L-1 of iodide,which is much lower than that of Ag-ISE.
中图分类号 O657.1
所属栏目 工作简报
基金项目
收稿日期 2005/2/24
修改稿日期
网络出版日期
作者单位点击查看
备注王建雅(1963-),女,辽宁省海城市人,教授,博士,从事电分析化学和光谱分析化学研究工作.
引用该论文: WANG Jian-ya,XUE Bin,YU Shi-jun,ZHAO Li-jie,BAI Xiao-lin. APPLICATION OF PVC-MEMBRANE IODIDE ION SELECTIVE ELECTRODE TO THE ANALYSIS OF SYSTEM CO-EXISTING OF BROMIDE AND IODIDE IONS[J]. Physical Testing and Chemical Analysis part B:Chemical Analysis, 2006, 42(12): 1010~1012
王建雅,薛斌,于士君,赵丽杰,白晓琳. 聚氯乙烯膜碘离子选择电极在溴碘离子共存体系分析中的应用[J]. 理化检验-化学分册, 2006, 42(12): 1010~1012
被引情况:
【1】王瑞侠,周享春,陆光汉, "电化学分析方法测定卤素的研究进展",理化检验-化学分册 48, 1123-1128(2012)
共有人对该论文发表了看法,其中:
人认为该论文很差
人认为该论文较差
人认为该论文一般
人认为该论文较好
人认为该论文很好
参考文献
【1】Ahmad R,Mohammad G,Mojtaba S.Lead ion selective PVC membrane electrode based on 5,5′-dithiobis-(2-nitrobenzoic acid)[J].Talanta,1998,46(6):1341-1346.
【2】Gholivand M B,Mozaffari Y.PVC-based bis (2-nitrophenyl) disulfide sensor for zinc ions[J].Talanta,2003,59(2):399-407.
【3】许宏鼎,王彦.聚氯乙烯膜对硝基苯酚离子选择电极的研制[J].分析化学,1991,19:1162-1164.
【4】王彦.PVC膜碘离子选择电极的制备及对食盐中碘含量的测定[J].分析化学,1996,24:863.
【5】王彦.PVC膜碘离子选择电极的研制[J].化学传感器,1991,11:60-64.
【2】Gholivand M B,Mozaffari Y.PVC-based bis (2-nitrophenyl) disulfide sensor for zinc ions[J].Talanta,2003,59(2):399-407.
【3】许宏鼎,王彦.聚氯乙烯膜对硝基苯酚离子选择电极的研制[J].分析化学,1991,19:1162-1164.
【4】王彦.PVC膜碘离子选择电极的制备及对食盐中碘含量的测定[J].分析化学,1996,24:863.
【5】王彦.PVC膜碘离子选择电极的研制[J].化学传感器,1991,11:60-64.
相关信息