Choice of Working Condition of Instrument in HG-AFS Determination of Trace Amounts of Arsenic
摘 要
作者等指出在氢化物发生-原子荧光光谱法的实施中,应注意在采用不同型号的仪器时选择专用的仪器工作条件,以免造成误差.文中叙述了应用AFS-820双通道原子荧光仪测定痕量砷时,对仪器的工作条件(包括灯电流及负高压)及样品溶液的化学条件(包括酸度及各种试剂的加入浓度)作了研究和优化.在所选择的条件下,方法的检出限为0.34 μg·L-1.在40 μg·L-1浓度水平上对测定砷的精密度进行试验,根据11次测定所得荧光强度的数据计算得相对标准偏差为1.4%.
Abstract
It was shown that in the practice of HG-AFS,specialized or exclusive instrumental working conditions should be chosen for different types or models of the atomic fluorescence spectrophotometer.It was reported that in the determination of trace amount of arsenic by HG-AFS using the AFS-820 Bi-channel atomic fluorescence spectrophotometer,working conditions of the instrument (including lamp current and negative high voltage) and chemical conditions of sample solution (including acidity and concentration of reagents added) were studied and optimized.Under the optimum conditions,detection limit of the method was found to be 0.34 μg·L-1.Precision was tested at the concentration level of 40 μg·L-1 with 11 determinations,values of RSD′s calculated from the data of fluorescence intensity measured was 1.4%.
中图分类号 O657.31
所属栏目 工作简报
基金项目
收稿日期 2007/12/3
修改稿日期
网络出版日期
作者单位点击查看
备注李学文(1978-),男,山东潍坊人,博士,主要研究方向为环境科学.
引用该论文: LI Xue-wen,JIN Lan-shu,LI Hui-jie,Lin Guo-lin. Choice of Working Condition of Instrument in HG-AFS Determination of Trace Amounts of Arsenic[J]. Physical Testing and Chemical Analysis part B:Chemical Analysis, 2009, 45(1): 73~74
李学文,金兰淑,李会杰,林国林. 氢化物发生-原子荧光光谱法测定痕量砷时仪器工作条件的选择[J]. 理化检验-化学分册, 2009, 45(1): 73~74
被引情况:
【1】吴庆晖,黄伯熹,杨运云,荀合,霍柱健, "氢化物发生-原子荧光光谱法测定六神丸及其在人工胃液中的可溶性砷含量",理化检验-化学分册 46, 885-887(2010)
【2】赵昌平, "氢化物发生-原子荧光光谱法同时测定环境水中砷和汞",理化检验-化学分册 47, 1452-1454(2011)
共有人对该论文发表了看法,其中:
人认为该论文很差
人认为该论文较差
人认为该论文一般
人认为该论文较好
人认为该论文很好
参考文献
【1】齐素芬.AFS-2202双道原子荧光计测定砷锡铋汞中应该注意的几个问题[J].岩矿测试,2006,25(2):197-198.
【2】袁爱萍,唐艳霞,黄玉龙,等.氢化物-原子荧光法测定锑精矿中痕量砷、汞的研究[J].光谱学与光谱分析,2006,26(8):1553-1556.
【3】陈海东,赵质创,孙亮,等.原子荧光光谱法同时测定农产品中砷和汞[J].理化检验-化学分册,2006,42(2):113-114.
【4】尹雪斌,姚春霞,骆永明,等.抽风口与烟道间距对AFS-930型原子荧光光度计测定砷、汞的影响[J].分析试验室,2006,25(10):119-122.
【5】王凯,彭珊茁,张杰,等.氢化物发生-原子荧光光谱法测定人发中砷含量[J].中国工业医学杂志,2005,18(6):331-333.
【6】刘晶,郑楚光,徐杰英.微波消解-冷原子荧光法测定煤中痕量汞[J].光谱学与光谱分析,2004,24(9):1113-1135.
【7】邬春华,吕元琦,袁倬斌.微波消解原子荧光光谱法测定生物样品中砷汞[J].理化检验-化学分册,2006,42(1):41-45.
【8】邹菁,喻德忠,刘安昌.液体硅肥中砷形态的氢化物发生原子荧光光谱分析[J].分析科学学报,2006,22(5):591-593.
【2】袁爱萍,唐艳霞,黄玉龙,等.氢化物-原子荧光法测定锑精矿中痕量砷、汞的研究[J].光谱学与光谱分析,2006,26(8):1553-1556.
【3】陈海东,赵质创,孙亮,等.原子荧光光谱法同时测定农产品中砷和汞[J].理化检验-化学分册,2006,42(2):113-114.
【4】尹雪斌,姚春霞,骆永明,等.抽风口与烟道间距对AFS-930型原子荧光光度计测定砷、汞的影响[J].分析试验室,2006,25(10):119-122.
【5】王凯,彭珊茁,张杰,等.氢化物发生-原子荧光光谱法测定人发中砷含量[J].中国工业医学杂志,2005,18(6):331-333.
【6】刘晶,郑楚光,徐杰英.微波消解-冷原子荧光法测定煤中痕量汞[J].光谱学与光谱分析,2004,24(9):1113-1135.
【7】邬春华,吕元琦,袁倬斌.微波消解原子荧光光谱法测定生物样品中砷汞[J].理化检验-化学分册,2006,42(1):41-45.
【8】邹菁,喻德忠,刘安昌.液体硅肥中砷形态的氢化物发生原子荧光光谱分析[J].分析科学学报,2006,22(5):591-593.
相关信息