FI-HG-CV-AAS Determination of Mercury(Ⅱ) Waste Water
摘 要
提出采用流动注射-氢化物发生-冷原子吸收光谱法测定废水中汞(Ⅱ),在优化的试验条件和合适的样品酸度及硼氢化钾浓度存在下,汞的检出限(3S/N)为0.006 μg·L-1,并用标准加入法作回收试验,得出回收率在94.0%~101.0%之间,相对标准偏差(n=6)在3.86%~4.61%之间.
Abstract
The FI-hydride generator was used in combination with the atomic absorption spectrometer for cold vapor AAS determination of Hg(Ⅱ) in waste water. Under the optimum conditions of acidity of sample solution,concentration of KBH4 added and operation parameters of the instrument,value of detection limit (3S/N) of this method attained to 0.006 μg·L-1. Recovery of the method was tested by the standard addition method,and the results obtained were in the range of 94.0% to 101.0%,with values of RSD′s (n=6) ranged from 3.9% to 4.7%.
中图分类号 O657.3
所属栏目 工作简报
基金项目
收稿日期 2009/4/7
修改稿日期
网络出版日期
作者单位点击查看
备注谢林灵(1973-),女,海南人,工程师,主要从事气相色谱、材料化学检测工作.
引用该论文: XIE Lin-ling. FI-HG-CV-AAS Determination of Mercury(Ⅱ) Waste Water[J]. Physical Testing and Chemical Analysis part B:Chemical Analysis, 2009, 45(7): 810~811
谢林灵. 流动注射-氢化物发生-冷原子吸收光谱法测定废水中汞(Ⅱ)[J]. 理化检验-化学分册, 2009, 45(7): 810~811