Cause Analysis on Fracture of Nickel Titanium Wires
摘 要
某镍钛合金植入类医疗器械产品里的镍钛丝在生产过程中发生断裂,采用化学成分分析、力学性能试验、洁净度测试、断口宏微观分析以及能谱分析等方法对其断裂原因进行了分析。结果表明:镍钛丝的加工应力未得到很好的消除,加之编织模棒和热缩管引入了腐蚀介质,二者共同作用导致该镍钛丝发生了应力腐蚀断裂。
Abstract
The nickel titanium wires in a nickel titanium alloy implantable medical device product fractured during production. The fracture reasons were analyzed by chemical composition analysis, mechanical property test, cleanliness test, macro and micro fracture analysis, energy spectrum analysis and so on. The results show that the machining stress of the nickel titanium wires had not been eliminated well, and the corrosive medium introduced by the braiding rod and the heat shrinkable tube. These two factors together caused the stress corrosion fracture of the nickel titanium wires.