Resonance Scattering Spectrometric Determination of Trace Amounts of Tin(Ⅱ) in Environmental Water Samples
摘 要
在pH 4.3的缓冲介质中,痕量锡(Ⅱ)对氯酚红-人血白蛋白-十二烷基硫酸钠体系的共振散射光谱有明显的减弱作用,共振散射强度的减弱程度(ΔI)与锡(Ⅱ)的质量浓度之间在0.2~4.0 μg·L-1范围内呈线性关系,检出限(3S/N)为0.12 μg·L-1,用于3件环境水样中痕量锡(Ⅱ)的测定,测定结果的相对标准偏差(n=6)小于3.5%,加标回收率为95.5%~100.6%.
Abstract
In a buffer medium of pH 4.3,intensity of resonance scattering of the reaction system of chlorophenol red (CPR),human serum albumin (HSA) and sodium dodecyl sulfate (SDS) was significantly decreased by the presence of stannous tin.Linear relationship was kept between the magnitude of decrease in intensity of resonance scattering (ΔI) and concentration of Tin(Ⅱ) in the range of 0.2-4.0 μg·L-1,with the detection limit (3S/N) of 0.12 μg·L-1.In its application to the analysis of 3 water samples,values of RSD′s (n=6) obtained were less than 3.5%.Results of recovery test found by standard addition method were in the range of 95.5%-100.6%.
中图分类号 O657.31
所属栏目 工作简报
基金项目
收稿日期 2007/5/17
修改稿日期
网络出版日期
作者单位点击查看
备注李少旦(1975-),男,邵阳市人,硕士,研究方向为有毒有害物质检测方法及分子机理的光谱学研究.
引用该论文: LI Shao-dan,LI Yan-zhi,LI Mu-lan. Resonance Scattering Spectrometric Determination of Trace Amounts of Tin(Ⅱ) in Environmental Water Samples[J]. Physical Testing and Chemical Analysis part B:Chemical Analysis, 2008, 44(10): 994~995
李少旦,李延志,李木兰. 共振散射光谱法测定环境水样中痕量锡(Ⅱ)[J]. 理化检验-化学分册, 2008, 44(10): 994~995
共有人对该论文发表了看法,其中:
人认为该论文很差
人认为该论文较差
人认为该论文一般
人认为该论文较好
人认为该论文很好
参考文献
【1】高建华,王弋,陈彬,等.氢化物发生微分电位溶出法测定人尿头发及自来水中痕量锡[J].理化检验-化学分册,2001,37(6):255-257,260.
【2】周之荣,周利民,王黎.偶氮胂-M催化动力学光度法测定痕量锡[J].稀有金属,2004,28(6):1109-1112.
【3】周之荣,熊艳.高碘酸钾氧化变色酸2R褪色光度法测定痕量锡Ⅳ[J].冶金分析,2005,25(6):17-21.
【4】张丽娟,金伟,杜光宇.车间空气中锡的氢化物发生-原子荧光法测定法[J].职业与健康,2004,20(4):45-47.