Advanced Search
    MAO Ai-rong. FAAS Determination of Copper in Chromium Plating Solution[J]. PHYSICAL TESTING AND CHEMICAL ANALYSIS PART B:CHEMICAL ANALYSIS, 2009, 45(7): 842-843.
    Citation: MAO Ai-rong. FAAS Determination of Copper in Chromium Plating Solution[J]. PHYSICAL TESTING AND CHEMICAL ANALYSIS PART B:CHEMICAL ANALYSIS, 2009, 45(7): 842-843.

    FAAS Determination of Copper in Chromium Plating Solution

    More Information
    • Received Date: April 10, 2008
    • FAAS was applied to the determination of copper in chromium plating solution. The chromium plating solution was acidified with conc. H2SO4,and then Cr(Ⅵ) was reduced into Cr(Ⅲ) by the addition of H2O2. Copper(Ⅱ) was precipitated completely by displacement with Zn from the sample solution,and dissolved in HNO3(1+1) and made up to 1 L with water,which was used for FAAS determination. The working condition of instrument was described in detail. Linear relationship between values of absorbance and mass concentration of Cu found was in the range within 4.0 mg·L-1,with values of RSD′s (n=6) 0.28%. Detection limit (3S/N) of the method obtained were 0.006 1 mg·L-1 respectively. The proposed method was applied to the analysis of 4 samples of chromium plating solutions,and value of recovery found by the standard addition method was in range of 95.0% to 106.0%.
    • [1]
      徐盘明,赵祥大.实用金属材料分析方法[M].安徽:中国科学技术大学出版社,1990:713-717.
      [2]
      GB/T 5009.13-2003食品中铜的测定方法[S].
      [3]
      孙淑清,曾铃,代丽双,等.火焰原子吸收光谱法测定葡萄中的铜含量[J].辽宁工学院学报,2005,25(6):419-420.
      [4]
      林穗云.镀铬液中杂质铜、锌的分析[J].嘉应大学学报,2000(3):65-66.

    Catalog

      Article views (3) PDF downloads (0) Cited by()

      /

      DownLoad:  Full-Size Img  PowerPoint
      Return
      Return